dc.description.resumo |
Um sistema de bombeamento de fluxo pulsátil permite replicar as formas de onda de
fluxo hidráulico, com precisão razoável para diversas simulações, como por exemplos fluxos
sanguíneos fisiológicos, fluxos hidráulicos de sistemas de baixa pressão, etc. Uma aplicação
importante desse sistema é permitir a simulação do fluxo sanguíneo do corpo humano
permitindo a análise da mecânica dos fluidos do sistema circulatório humano, com um
controle de variáveis como pressão, temperatura e ciclos por minuto, podendo-se ter uma
condição muito próxima da realidade, além de permitir criar diversas situações distintas
modificando os valores das variáveis mencionadas anteriormente. Para análise de Stents
cardiovasculares, por exemplo, a Norma ASTM 2477-7 recomenda que os testes sejam feitos
em condições semelhantes aquelas de uso, ou seja, em fluxo pulsátil simulando o coração
humano. Stent é um dispositivo médico que pode ser definido como um cilindro de metal ou
polímero, geralmente em forma de rede, que é colocado no lúmen de um vaso anatômico
(como uma artéria ou um duto biliar) com o objetivo de reabrir a passagem desse vaso que
encontra-se total ou parcialmente obstruída, geralmente com o intuito de manter um normal
fluxo do sangue. Em virtude de sua aplicação influenciar diretamente a saúde do paciente,
tem-se à necessidade de pesquisar formas de melhorar as propriedades físicas e químicas,
tornando a tecnologia dos Stents mais confiável sem comprometer a biocompatibilidade.
Assim, o presente trabalho, propõe o desenvolvimento de uma bancada experimental de fluxo
pulsátil que poderá ser usada para ensaios em Stents. O projeto foi dividido em duas partes, a
primeira foi o planejamento e fabricação da bancada (Projeto conceitual e mecânico) e em um
segundo momento, a instrumentação da mesma (Projeto eletrônico). Como subsídio do
projeto foram estudadas diversas bancadas geradoras fluxo pulsátil (que simulam o coração
humano) e a partir desses, concebida e fabricada uma bancada que trabalha com fluido
escoando em ciclo fechado. A mesma é composta por um atuador linear (Gerador de Pulso) e
vários elementos sensores que permitem o sistema trabalhar em malha fechada, com controle
de temperatura, pressão e número de ciclos por minuto. Todos os dados gerados são
armazenados em um banco de dados podendo ser monitorados remotamente via web. Para
análise, o corpo de prova é colocado no interior do tudo ou em uma célula de ensaio e um
fluido, é circulado de forma pulsátil pelas mangueiras. |
pt_BR |
dc.publisher.country |
Brasil |
pt_BR |
dc.publisher.department |
Centro de Ciências e Tecnologia - CCT |
pt_BR |
dc.publisher.program |
PÓS-GRADUAÇÃO EM ENGENHARIA MECÂNICA |
pt_BR |
dc.publisher.initials |
UFCG |
pt_BR |
dc.subject.cnpq |
Engenharia Mecânica. |
pt_BR |
dc.title |
Projeto, fabricação e testes de uma bancada experimental de fluxo pulsátil. |
pt_BR |
dc.date.issued |
2015-08-28 |
|
dc.description.abstract |
A pulsatile flow pumping system allows replicating forms hydraulic flow wave, with
reasonable accuracy for various simulations, for example physiological blood flows, inflows
of low pressure hydraulic systems, etc. An important application of this system is to enable
simulation of blood flow in the human body allowing the analysis of the mechanics of the
human circulatory system fluids, with a control variable such as pressure, temperature and
cycles per minute, being able to have a very next condition reality, and allows creating many
different situations modifying the values of the variables mentioned above. For analysis of
cardiovascular stents, for example, ASTM 2477-7 recommended that tests be conducted under
conditions similar to those of use, ie in pulsatile flow simulating human heart. The stent is a
medical device that can be defined as a metal cylinder or polymer, usually in network form,
which is placed in the lumen of an anatomical vessel (such as an artery or a bile duct) in order
to reopen the passage of this vessel which is fully or partially obstructed, usually in order to
maintain a normal blood flow. Because of its application directly influence the health of the
patient, there is the need to search for ways to improve the physical and chemical properties,
making the most reliable technology of stents without sacrificing the biocompatibility. Thus,
this paper proposes the development of a pulsatile flow test rig that can be used for testing on
stents. The project was divided into two stages, the first was the design and craftsmanship of
the bench (Conceptual Design and Mechanical) and in a second step, the same
instrumentation (electronic design). As project grant were studied various pulsatile flow
generating benches (which simulate the human heart) and from the these, designed and
manufactured a bench working with leaking fluid in a closed cycle. The same consists of a
linear actuator (Pulse Generator) and several sensor elements that allow the system to work in
closed loop, with temperature control, pressure and number of cycles per minute. All data
generated is stored in a database and can be monitored remotely via the web. For analysis, the
test body is placed inside all or a test cell, and a fluid is circulated through the tubes pulsatile. |
pt_BR |
dc.identifier.uri |
http://dspace.sti.ufcg.edu.br:8080/jspui/handle/riufcg/28257 |
|
dc.date.accessioned |
2022-12-09T21:53:35Z |
|
dc.date.available |
2022-12-09 |
|
dc.date.available |
2022-12-09T21:53:35Z |
|
dc.type |
Dissertação |
pt_BR |
dc.subject |
Bancada experimental de fluxo pulsátil |
pt_BR |
dc.subject |
Fluxo pulsátil |
pt_BR |
dc.subject |
Fluxo em malha fechada |
pt_BR |
dc.subject |
Stents cardiovasculares |
pt_BR |
dc.subject |
Norma ASTM 2477-07 |
pt_BR |
dc.subject |
Ciclo cardíaco |
pt_BR |
dc.subject |
Sistema de bombeamento de fluxo pulsátil |
pt_BR |
dc.subject |
Experimental pulsatile flow bench |
pt_BR |
dc.subject |
Pulsatile flow |
pt_BR |
dc.subject |
Closed loop flow |
pt_BR |
dc.subject |
Cardiovascular stents |
pt_BR |
dc.subject |
ASTM Standard 2477-07 |
pt_BR |
dc.subject |
Cardiac cycle |
pt_BR |
dc.subject |
Pulsed flow pumping system |
pt_BR |
dc.rights |
Acesso Aberto |
pt_BR |
dc.creator |
JERONIMO, Emerson. |
|
dc.publisher |
Universidade Federal de Campina Grande |
pt_BR |
dc.language |
por |
pt_BR |
dc.title.alternative |
Design, fabrication and testing of an experimental pulsatile flow bench. |
pt_BR |
dc.identifier.citation |
JERONIMO, Emerson. Projeto, fabricação e testes de uma bancada experimental de fluxo pulsátil. 2015. 106f. (Dissertação de Mestrado) Programa de Pós-Graduação em Engenharia Mecânica, Centro de Ciências e Tecnologia, Universidade Federal de Campina Grande - Paraíba - Brasil, 2015. Disponível em: http://dspace.sti.ufcg.edu.br:8080/jspui/handle/riufcg/28257 |