DSpace/Manakin Repository

Irradiância dos aparelhos de fototerapia em uma maternidade de João Pessoa.

Mostrar registro simples

dc.creator.ID ALVES, L. F. pt_BR
dc.creator.ID CATUNDA. S. B.
dc.creator.Lattes http://lattes.cnpq.br/8615902149888988 pt_BR
dc.creator.Lattes http://lattes.cnpq.br/4942491863397997
dc.contributor.advisor1 AZEVEDO, Waldeneide Fernandes de.
dc.contributor.advisor1ID AZEVEDO, W. F. pt_BR
dc.contributor.advisor1Lattes http://lattes.cnpq.br/6917079139731942 pt_BR
dc.contributor.referee2 TRINDADE, Mônica Cavalcanti.
dc.contributor.referee2ID TRINDADE, M. C. pt_BR
dc.description.resumo INTRODUÇÂO: A fototerapia tem sido a conduta universalmente mais difundida para o tratamento da icterícia neonatal. O correto manejo, bem como aparelhos e lâmpadas adequadas com contínuo monitoramento de suas irradiâncias, são de fundamental importância para que a fototerapia seja eficaz no tratamento da doença, reduzindo o tempo de internamento e o risco de dano cerebral. OBJETIVO: Avaliar a eficácia dos aparelhos de fototerapia utilizados em recém-nascidos no Instituto Cândida Vargas (ICV) através da medição padronizada da irradiância. METODOLOGIA: Trata-se de um estudo transversal com caráter descritivo e abordagem quantitativa. A amostra foi composta por todos os aparelhos de fototerapia disponíveis no serviço, totalizando 17. Para aferição foi utilizado um fotodosímetro modelo THOR Multitester 3620, calibrado diretamente em microwatts por centímetros ao quadrado por nanômetros (μw/cm2/nm). Os dados foram tabulados em planilha Excel e analisados descritivamente. RESULTADO: Foram analisados 17 aparelhos de fototerapia, sendo 16 Bilispot® e 1 Biliberço®. A irradiância média dos aparelhos foi de 2,95 μW/cm²/nm ± 1,42. Os aparelhos do tipo Bilispot® apresentaram uma variação de 0,97 a 5,32 μW/cm²/nm, média de 2,77 μW/cm²/nm; o Biliberço® apresentou a maior irradiância entre todos os aparelhos analisados, com um valor de 5,87 μW/cm²/nm. Apenas 17,64% dos aparelhos apresentaram uma irradiância maior que 4 μW/cm²/nm. CONCLUSÃO: A maioria dos aparelhos estavam com irradiância inferior ao mínimo considerado terapêutico, tornando evidente a necessidade de uma maior eficácia na manutenção dos aparelhos de fototerapia. pt_BR
dc.publisher.country Brasil pt_BR
dc.publisher.department Centro de Ciências Biológicas e da Saúde - CCBS pt_BR
dc.publisher.initials UFCG pt_BR
dc.subject.cnpq Medicina. pt_BR
dc.title Irradiância dos aparelhos de fototerapia em uma maternidade de João Pessoa. pt_BR
dc.date.issued 2018-08-01
dc.description.abstract INTRODUCTION: Phototherapy has been universally most widely used way to treat neonatal jaundice, and the correct management, as well as adequate devices and lamps with continuous monitoring of their irradiances, are of fundamental importance for effective phototherapy in the treatment of jaundice, reducing the length of hospitalization and the risk of brain damage. OBJECTIVE: Evaluate the effectiveness of phototherapy devices used in newborns at Instituto Cândida Vargas (ICV) through the standardized measurement of the irradiance. STUDY DESIGN: This is a descriptive cross-sectional study with a quantitative approach. To measure the 17 phototherapy devices, a THOR Multitester 3620 photodosimeter was used, calibrated directly in microwatts per square centimeters per nanometer (mw / cm2 / nm). RESULTS: 17 phototherapy devices, 16 Bilispot® and 1 Biliberço®, were analyzed. The mean irradiance of the devices was 2.95 μW/cm²/nm ± 1.42. The Bilispot®-type devices presented a variation of 0.97 to 5.32 μW/cm²/nm, average of 2.77 μW/cm²/nm; Biliberço® presented the highest irradiance among all analyzed devices, with a value of 5.87 μW/cm²/nm. Only 17.64% of the devices had an irradiance greater than 4 μW/cm²/nm. CONCLUSION: The majority of the devices had irradiance below the minimum considered therapeutic, making evident the need for a greater efficiency in the maintenance of phototherapy devices. pt_BR
dc.identifier.uri http://dspace.sti.ufcg.edu.br:8080/jspui/handle/riufcg/30625
dc.date.accessioned 2023-07-09T20:37:01Z
dc.date.available 2023-07-09
dc.date.available 2023-07-09T20:37:01Z
dc.type Trabalho de Conclusão de Curso pt_BR
dc.subject Icterícia pt_BR
dc.subject Fototerapia pt_BR
dc.subject Icterícia neonatal pt_BR
dc.subject Jaundice pt_BR
dc.subject Phototherapy pt_BR
dc.subject Neonatal jaundice pt_BR
dc.subject Estudo transversal com caráter descritivo e abordagem quantitativa pt_BR
dc.subject Irradiância - equipamentos hospitalares - maternidade pt_BR
dc.subject Maternidade - João Pessoa - PB pt_BR
dc.rights Acesso Aberto pt_BR
dc.creator ALVES, Lívia Feitosa.
dc.creator CATUNDA, Stephanie Bento.
dc.publisher Universidade Federal de Campina Grande pt_BR
dc.language por pt_BR
dc.title.alternative Irradiance of phototherapy devices in a maternity hospital in João Pessoa. pt_BR
dc.identifier.citation ALVES, Lívia Feitosa. CATUNDA, Stephanie Bento. Irradiância dos aparelhos de fototerapia em uma maternidade de João Pessoa. 2018. 35fl. (Trabalho de Conclusão de Curso – Monografia), Curso de Bacharelado em Medicina, Centro de Ciências Biológicas e da Saúde, Universidade Federal de Campina Grande – Paraíba – Brasil, 2018. pt_BR