dc.description.resumo |
INTRODUÇÂO: A fototerapia tem sido a conduta universalmente mais difundida para o tratamento da icterícia neonatal. O correto manejo, bem como aparelhos e lâmpadas adequadas com contínuo monitoramento de suas irradiâncias, são de fundamental importância para que a fototerapia seja eficaz no tratamento da doença, reduzindo o tempo de internamento e o risco de dano cerebral. OBJETIVO: Avaliar a eficácia dos aparelhos de fototerapia utilizados em recém-nascidos no Instituto Cândida Vargas (ICV) através da medição padronizada da irradiância.
METODOLOGIA: Trata-se de um estudo transversal com caráter descritivo e abordagem quantitativa. A amostra foi composta por todos os aparelhos de fototerapia disponíveis no serviço, totalizando 17. Para aferição foi utilizado um fotodosímetro modelo THOR Multitester 3620, calibrado diretamente em microwatts por centímetros ao quadrado por nanômetros (μw/cm2/nm). Os dados foram tabulados em planilha Excel e analisados descritivamente. RESULTADO: Foram analisados 17 aparelhos de fototerapia, sendo 16 Bilispot® e 1 Biliberço®. A irradiância média dos aparelhos foi de 2,95 μW/cm²/nm ± 1,42. Os aparelhos do tipo Bilispot® apresentaram uma variação de 0,97 a 5,32 μW/cm²/nm, média de 2,77 μW/cm²/nm; o Biliberço® apresentou a maior irradiância entre todos os aparelhos analisados, com um valor de 5,87 μW/cm²/nm. Apenas 17,64% dos aparelhos apresentaram uma irradiância maior que 4 μW/cm²/nm. CONCLUSÃO: A maioria dos aparelhos estavam com irradiância inferior ao mínimo considerado terapêutico, tornando evidente a necessidade de uma maior eficácia na
manutenção dos aparelhos de fototerapia. |
pt_BR |
dc.publisher.country |
Brasil |
pt_BR |
dc.publisher.department |
Centro de Ciências Biológicas e da Saúde - CCBS |
pt_BR |
dc.publisher.initials |
UFCG |
pt_BR |
dc.subject.cnpq |
Medicina. |
pt_BR |
dc.title |
Irradiância dos aparelhos de fototerapia em uma maternidade de João Pessoa. |
pt_BR |
dc.date.issued |
2018-08-01 |
|
dc.description.abstract |
INTRODUCTION: Phototherapy has been universally most widely used way to treat
neonatal jaundice, and the correct management, as well as adequate devices and
lamps with continuous monitoring of their irradiances, are of fundamental
importance for effective phototherapy in the treatment of jaundice, reducing the
length of hospitalization and the risk of brain damage. OBJECTIVE: Evaluate the
effectiveness of phototherapy devices used in newborns at Instituto Cândida Vargas
(ICV) through the standardized measurement of the irradiance. STUDY DESIGN:
This is a descriptive cross-sectional study with a quantitative approach. To measure
the 17 phototherapy devices, a THOR Multitester 3620 photodosimeter was used,
calibrated directly in microwatts per square centimeters per nanometer (mw / cm2 /
nm). RESULTS: 17 phototherapy devices, 16 Bilispot® and 1 Biliberço®, were
analyzed. The mean irradiance of the devices was 2.95 μW/cm²/nm ± 1.42. The
Bilispot®-type devices presented a variation of 0.97 to 5.32 μW/cm²/nm, average of
2.77 μW/cm²/nm; Biliberço® presented the highest irradiance among all analyzed
devices, with a value of 5.87 μW/cm²/nm. Only 17.64% of the devices had an
irradiance greater than 4 μW/cm²/nm. CONCLUSION: The majority of the devices
had irradiance below the minimum considered therapeutic, making evident the need
for a greater efficiency in the maintenance of phototherapy devices. |
pt_BR |
dc.identifier.uri |
http://dspace.sti.ufcg.edu.br:8080/jspui/handle/riufcg/30625 |
|
dc.date.accessioned |
2023-07-09T20:37:01Z |
|
dc.date.available |
2023-07-09 |
|
dc.date.available |
2023-07-09T20:37:01Z |
|
dc.type |
Trabalho de Conclusão de Curso |
pt_BR |
dc.subject |
Icterícia |
pt_BR |
dc.subject |
Fototerapia |
pt_BR |
dc.subject |
Icterícia neonatal |
pt_BR |
dc.subject |
Jaundice |
pt_BR |
dc.subject |
Phototherapy |
pt_BR |
dc.subject |
Neonatal jaundice |
pt_BR |
dc.subject |
Estudo transversal com caráter descritivo e abordagem quantitativa |
pt_BR |
dc.subject |
Irradiância - equipamentos hospitalares - maternidade |
pt_BR |
dc.subject |
Maternidade - João Pessoa - PB |
pt_BR |
dc.rights |
Acesso Aberto |
pt_BR |
dc.creator |
ALVES, Lívia Feitosa. |
|
dc.creator |
CATUNDA, Stephanie Bento. |
|
dc.publisher |
Universidade Federal de Campina Grande |
pt_BR |
dc.language |
por |
pt_BR |
dc.title.alternative |
Irradiance of phototherapy devices in a maternity hospital in João Pessoa. |
pt_BR |
dc.identifier.citation |
ALVES, Lívia Feitosa. CATUNDA, Stephanie Bento. Irradiância dos aparelhos de fototerapia em uma maternidade de João Pessoa. 2018. 35fl. (Trabalho de Conclusão de Curso – Monografia), Curso de Bacharelado em Medicina, Centro de Ciências Biológicas e da Saúde, Universidade Federal de Campina Grande – Paraíba – Brasil, 2018. |
pt_BR |