Dépôt DSpace/Manakin

Estudo da degradação de ponteiras de endoscópios utilizadas em endoscopia digestiva alta.

Afficher la notice abrégée

dc.contributor.advisor1 MENEZES, Romualdo Rodrigues.
dc.contributor.advisor1ID MENEZES, R. R. pt_BR
dc.contributor.advisor1Lattes http://lattes.cnpq.br/1169292996645931 pt_BR
dc.contributor.advisor2 FOOK, Marcus Vinícius Lia.
dc.contributor.advisor2ID FOOK, M. V. L. pt_BR
dc.contributor.advisor2Lattes http://lattes.cnpq.br/4149843752530120 pt_BR
dc.contributor.referee1 BARBOSA, Rossemberg Cardoso.
dc.contributor.referee2 LIMA, Rodrigo José da Silva.
dc.description.resumo O aparelho endoscópico utilizado atualmente para estudo do trato gastro-intestinal (o video-endoscópio) é resultado de quase 200 anos de evolução tecnológica. Estes endoscópios não podem ser esterilizados pelos métodos clássicos. Por sua complexidade, o processo de limpeza e desinfecção de endoscópios não é somente uma preocupação da enfermagem, ela se tornou multidisciplinar, pois exige discussão sobre riscos físicos, biológicos e químicos, tanto para o paciente quanto para a equipe. A borracha da ponta flexível dos endoscópios é uma peça fundamental para a preservação da vedação do endoscópio e principalmente preservação do paciente. Esta borracha sofre desgaste tanto durante o exame como pelos processos de desinfecção do equipamento que acabam agredindo a superfície do material. O seu desgaste ou ruptura pode acarretar na queima do gerador de imagem. A troca da borracha da ponta é considerada uma manutenção parcial e deve sempre ser executada periodicamente. Baseado no exposto este trabalho teve como objetivo principal estudar e avaliar a degradação ocorrida em ponteiras flexíveis de endoscópios digestivos submetidos a procedimentos médicos. Como resultados pode-se verificar nas microscopia, tanto ótica como eletrônica, morfologia rugosa com presença de poros para a ponteira virgem, nas ponteiras utilizadas por diferentes períodos de tempo verificou-se intensificação destas rugosidades. Por DRX pode-se observar um perfil praticamente amorfo típico de elastômero, além de identificar a presença de compostos de carga no polímero. Por FTIR pode-se comprovar diferentes estruturas para as amostras estudadas, levando a crer que algumas possuem polidimetilsiloxana em excesso com relação ao fluorelastômero. Por EDS pode-se verificar a exteriorização de compostos presentes no interior do elastômero após o período de degradação. No ensaio de tração foi verificada diminuição das propriedades físicas do elastômero devido a degradação ocorrida. No ensaio de ângulo de molhabilidade verificou-se que mesmo após o processo de degradação todas as amostras ainda apresentavam-se hidrofóbicas. Em síntese, este trabalho observou a necessidade de se desenvolver um material específico para utilização como ponteiras de endoscópios, já que os existentes no mercado aparentemente não tem destino especifico para aplicações biológicas. Ou seja, mesmo em sua forma virgem o elastômero em questão já apresenta fatores que não tornam sua utilização em meio biológico tão viável. pt_BR
dc.publisher.country Brasil pt_BR
dc.publisher.department Centro de Ciências e Tecnologia - CCT pt_BR
dc.publisher.program PÓS-GRADUAÇÃO EM CIÊNCIA E ENGENHARIA DE MATERIAIS pt_BR
dc.publisher.initials UFCG pt_BR
dc.subject.cnpq Ciência e Engenharia de Materiais. pt_BR
dc.title Estudo da degradação de ponteiras de endoscópios utilizadas em endoscopia digestiva alta. pt_BR
dc.date.issued 2014-12-18
dc.description.abstract The endoscopic device currently used to study the gastrointestinal tract (the videoendoscope) is the result of nearly 200 years of technological evolution. These endoscopes can not be sterilized by the classical methods. Due to its complexity, the process of cleaning and disinfection of endoscopes is not only a concern of nursing, she became multidisciplinary, since it requires discussion of physical, biological and chemical hazards, both for the patient and for the team. The flexible rubber tip of the endoscope is a key piece to preserve the seal of the endoscope and especially preservation of the patient. This rubber wears out both during the test as the disinfection of equipment that processes attacking the end surface of the material. Your wear or breakage may result in the image generator burns. The replacement of the rubber tip is considered a partial maintenance and must always be performed periodically. Based on the above this work aimed to study and evaluate the degradation occurred in flexible tips of digestive endoscopes undergoing medical procedures. The results can be seen in microscopy, both optical and electronic, rough morphology with the presence of pores to the virgin tip, the tips used for different periods of time there was intensification of these ridges. XRD can observe typical listing a substantially amorphous elastomer, and identify the presence of compounds in the polymer charge. FTIR can prove different structures for the studied samples, leading to the belief that some have polydimethylsiloxane in excess with respect to the fluoroelastomer. EDS can verify the externalization of compounds within the elastomer after the degradation period. In the tensile test was observed decrease in physical properties of the elastomer due to degradation occurred. On test wetting angle was found that even after the degradation process all the samples still had become hydrophobic. In summary, this study observed the need to develop a specific material for use as tips of endoscopes, since on the market apparently has no specific target for biological applications. That is, even in its virgin form the elastomer in question already has factors that do not make their use in biological environment as possible. pt_BR
dc.identifier.uri http://dspace.sti.ufcg.edu.br:8080/jspui/handle/riufcg/399
dc.date.accessioned 2018-04-17T12:25:43Z
dc.date.available 2018-04-17
dc.date.available 2018-04-17T12:25:43Z
dc.type Dissertação pt_BR
dc.subject Ponteiras de Endoscópios pt_BR
dc.subject Alta endoscopia digestiva pt_BR
dc.subject Fluorelastômetro pt_BR
dc.subject Video-endoscópio pt_BR
dc.subject High Digestive Endoscopy pt_BR
dc.subject Endoscopes Tips pt_BR
dc.subject Desgaste de endoscópios. pt_BR
dc.subject Wear of endoscopes. pt_BR
dc.subject Manutenção dos aparelhos de endoscopia pt_BR
dc.subject Microscopia Óptica (MO) pt_BR
dc.subject Microscopia Eletrônica de Varredura (MEV) pt_BR
dc.rights Acesso Aberto pt_BR
dc.creator ANDRADE, Antônio Augusto Fraga de.
dc.publisher Universidade Federal de Campina Grande pt_BR
dc.language por pt_BR
dc.title.alternative Study of the degradation of endoscopes tips used in upper digestive endoscopy. pt_BR
dc.identifier.citation ANDRADE, Antônio Augusto Fraga de. Estudo da degradação de ponteiras de endoscópios utilizadas em endoscopia digestiva alta. 2014. 65f . (Dissertação de Mestrado), Programa de Pós-graduação em Ciência e Engenharia de Materiais, Centro de Ciências e Tecnologia, Universidade Federal de Campina Grande - Paraíba - Brasil, 2014. Disponível em: http://dspace.sti.ufcg.edu.br:8080/jspui/handle/riufcg/399 pt_BR