dc.description.resumo |
A demanda por novos produtos aliada ao surgimento de novas técnicas de manufatura e à união
de ciências promissoras possibilita o desenvolvimento de dispositivos customizados para a
saúde feminina. Neste cenário, dispositivos de biofeedback mecânico que atuam como
adjuvantes no tratamento das disfunções do assoalho pélvico, desenvolvidos com características
customizadas utilizando novas tecnologias, podem fornecer informações durante o seu uso por
meio de estímulos visuais e sensoriais. Diante disso, esta pesquisa teve como objetivo
desenvolver um dispositivo mecânico para treinamento dos músculos do assoalho pélvico com
uso da manufatura aditiva. O estudo realizado foi de caráter científico de natureza aplicada,
com objetivo exploratório, de abordagem qualitativa, a partir do método de pesquisa-ação,
sendo dividido em seis etapas: análise biomimética, desenvolvimento conceptivo, modelagem
3D, simulação computacional, fabricação por manufatura aditiva, teste de bancada e análise do
protótipo por profissionais da saúde. Para o desenvolvimento conceitual e parametrização do
dispositivo adotou-se a bioinspiração no caju e no canal vaginal, sendo utilizados recursos como
o escaneamento 3D e imagens de TC para obtenção dos contornos, que serviram como base
para modelagem do dispositivo em softwares CAD. A partir da simulação computacional de
elementos finitos, o dispositivo foi aprovado para uso quanto à capacidade de permitir a
realização de contrações voluntárias do assoalho pélvico durante o treinamento dos músculos
do assoalho pélvico (TMAP). A fabricação feita por MA utilizando a tecnologia MJP 3 Polyjet
foi satisfatória para validar o modelo do design. Ao submeter o dispositivo a prática simulada
em modelo anatômico, constatou-se que os requisitos pretendidos na concepção inicial foram
atendidos, com destaque para a visibilidade do dispositivo pela paciente, durante o uso, em
posição de litotomia. O protótipo do dispositivo ao ser analisado por profissionais da saúde foi
bem avaliado e considerado apropriado para uso. Desta forma, o estudo mostrou-se satisfatório
em todas as suas fases, desde a concepção inicial até as etapas de modelagem, simulação e
fabricação. O desenvolvimento de um dispositivo mecânico para TMAP utilizando MA tem o
potencial de aprimorar o tratamento em ambiente clínico e facilitar a continuidade da terapia
pela paciente em domicílio, oferecendo uma abordagem integrada e eficaz para o cuidado com
a saúde. |
pt_BR |
dc.publisher.country |
Brasil |
pt_BR |
dc.publisher.department |
Centro de Ciências e Tecnologia - CCT |
pt_BR |
dc.publisher.program |
PÓS-GRADUAÇÃO EM ENGENHARIA DE PROCESSOS |
pt_BR |
dc.publisher.initials |
UFCG |
pt_BR |
dc.subject.cnpq |
Engenharia de Processos |
pt_BR |
dc.title |
Desenvolvimento de dispositivo para treinamento dos músculos do assoalho pélvico por manufatura aditiva |
pt_BR |
dc.date.issued |
2024-03-27 |
|
dc.description.abstract |
The demand for new products combined with the emergence of new manufacturing techniques
and the combination of promising sciences makes it possible to develop customized devices for
women's health. In this scenario, mechanical biofeedback devices that act as adjuvants in the
treatment of pelvic floor disorders, developed with customized characteristics using new
technologies, can provide information during their use through visual and sensory stimuli.
Therefore, this research aimed to develop a mechanical device for training the pelvic floor
muscles using additive manufacturing. The study carried out was of a scientific nature of an
applied nature, with an exploratory objective, with a qualitative approach, based on the action
research method, being divided into six stages: biomimetic analysis, conceptual development,
3D modeling, computer simulation, manufacturing by additive manufacturing, bench testing
and analysis of the prototype by healthcare professionals. For the conceptual development and
parameterization of the device, bioinspiration was adopted from cashews and the vaginal canal,
using resources such as 3D scanning and CT images to obtain the contours, which served as the
basis for modeling the device in CAD software. Based on computational finite element
simulation, the device was approved for use in terms of its ability to allow voluntary
contractions of the pelvic floor during pelvic floor muscle training (PFMT). The manufacturing
carried out by MA using MJP 3 Polyjet technology was satisfactory for validating the design
model. When submitting the device to simulated practice in an anatomical model, it was found
that the requirements intended in the initial design were met, with emphasis on the visibility of
the device by the patient, during use, in the lithotomy position. The device prototype, when
analyzed by health professionals, was well evaluated and considered appropriate for use. In this
way, the study proved to be satisfactory in all its phases, from the initial conception to the
modeling, simulation and manufacturing stages. The development of a mechanical device for
PFMT using AM has the potential to improve treatment in a clinical environment and facilitate
the continuation of therapy by the patient at home, offering an integrated and effective approach
to healthcare. |
pt_BR |
dc.identifier.uri |
http://dspace.sti.ufcg.edu.br:8080/jspui/handle/riufcg/41140 |
|
dc.date.accessioned |
2025-03-18T12:56:59Z |
|
dc.date.available |
2025-03-17 |
|
dc.date.available |
2025-03-18T12:56:59Z |
|
dc.type |
Tese |
pt_BR |
dc.subject |
Dispositivo médico |
pt_BR |
dc.subject |
Biomimética |
pt_BR |
dc.subject |
Simulação por computador |
pt_BR |
dc.subject |
Manufatura Aditiva |
pt_BR |
dc.subject |
Distúrbios do assoalho pélvico |
pt_BR |
dc.subject |
Equipment and Supplies |
pt_BR |
dc.subject |
Biomimetics |
pt_BR |
dc.subject |
Computer Simulation |
pt_BR |
dc.subject |
Additive manufacturing |
pt_BR |
dc.subject |
Pelvic floor disorders |
pt_BR |
dc.rights |
Acesso Aberto |
pt_BR |
dc.creator |
MARTINS, Ketinlly Yasmyne Nascimento. |
|
dc.publisher |
Universidade Federal de Campina Grande |
pt_BR |
dc.language |
por |
pt_BR |
dc.title.alternative |
Development of a device for training pelvic floor muscles by additive manufacturing |
pt_BR |
dc.description.sponsorship |
CNPq |
pt_BR |
dc.description.sponsorship |
Capes |
pt_BR |
dc.identifier.citation |
MARTINS, Ketinlly Yasmyne Nascimento. Desenvolvimento de dispositivo para treinamento dos músculos do assoalho pélvico por manufatura aditiva. 2024. 105 f. Tese (Doutorado em Engenharia de Processos) – Programa de Pós-Graduação em Engenharia de Processos, Centro de Ciências e Tecnologia, Universidade Federal de Campina Grande, Paraíba, Brasil,2024. |